Alarm Codes and Error Messages

Alarm Codes and Error Messages


ALARM MESSAGE
POSSIBLE CAUSE
POSSIBLE SOLUTION


High Temperature Alarm
The temperature has gone above the high temperature alarm point.
  1. Try to lower the temperature by turning heaters down or off, or by increasing or turning on cooling elements (such as fans or misters), or by a combination of both.
  2. Check the temperature probes.
  3. Check to see if a fan has failed.
  4. Check the alarm settings.


Low Temperature Alarm
The temperature has gone below the low temperature alarm point.
  1. Try to raise the temperature by turning heaters up or on, or by decreasing or turning off cooling elements (such as fans or misters), or by a combination of both.
  2. Check the temperature probes.
  3. Check to see if a heater has failed.
  4. Check the alarm settings. 

Probe Damage Alarm
A temperature probe is damaged or disconnected.
  1. Replace or reconnect the temperature probe. The control should recover automatically.
  2. If another probe is available, the control uses it. If probe averaging is on, the defective probe is excluded from the average.


Probe Deviation Alarm

The difference in measured temperatures between averaged probes is greater than 18°F (10°C).
  1. Check the temperature probes.

  2. Check to see if a cooling or heating element has failed.

  3. Check the alarm setting. 


Error 1 - SAVER Error
  1. The SEC-HD Saver is not in place when trying to save or restore settings.
  2. The Saver inserted is not an SEC-HD Saver
  3. The Saver inserted has invalid settings on it.
  1. Make sure the SEC-HD Saver is inserted correctly and then try again.
  2. Remove the Saver and then insert an SEC-HD Saver.
  3. Resave the settings from the SEC-HD to the SEC-HD Saver

Power Fail Alarm
There has been a power failure.
  1. Check to see why the power failed and then fix the problem.